Configuration
Selecting or disabling the capture ranges (only applies for the measuring
mode):
The capture range is selected in the measuring mode (factory setting) and
permanently disabled in adjustment mode.
The CC-Vision PC software can be used to select or disable the capture ranges
for the measuring mode.
5.1

Instrument settings

The following parameters can be changed on an instrument:
Name
Password
Operating signal LED
Horn operating signal
Switch-off mode
Shift length (TWA)
Short-term exposure level (STEL)
1)
At least one of the two operating signals must be switched on.
2)
Corresponds to the averaging time and is used to calculate the TWA exposure level.
3)
Interpretation only if the sensor is designed for this.
4)
Corresponds to the averaging time and is used to calculate the STEL exposure level.
5.2

Sensor settings

The following parameters can be changed on the sensors:
Name
Alarm threshold A1 (in measuring unit)
Alarm threshold A2 (in measuring unit)
Interpretation type
Alarm threshold STEL
(in measuring unit)
34
1)
1)
2)
3)4)
1)
1)
Range
Numerical range (3-digit)
Yes/No
Yes/No
"switch-off allowed" or
"switch-off prohibited" or
"switch-off prohibited at A2"
60 - 14400 (in minutes)
(Setting for exposure alarm)
0 - 15 (in minutes)
(setting for exposure alarm)
Range
Capture range – A2
A1 – measuring range end value
Inactive, TWA, STEL, TWA+STEL
Capture range – full scale value

Werbung

Inhaltsverzeichnis