Dräger X-am 5600 Gebrauchsanweisung Seite 47

Vorschau ausblenden Andere Handbücher für X-am 5600:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
8
Press the [OK] key to confirm the test gas concentration or use the [ + ] key
to change the test gas concentration and complete the process by pressing
the [OK] key.
The measurement value flashes.
9
Open the test gas cylinder valve to let gas flow over the sensor with a
volume flow of 0.5 L/min.
The displayed, flashing measurement value changes to the value according
to the supplied test gas.
When the displayed measurement value is stable (after at least 120 seconds):
Press the [OK] key to perform the adjustment.
a.
The display containing the current gas concentration changes with the
display » OK «.
Press the [OK] key or wait for approx. 5 seconds to end the adjustment
b.
of this measuring channel.
The next measuring channel is displayed for adjustment if necessary.
After the adjustment of the last measuring channel, the instrument
changes to the measuring mode.
c.
Close the test gas cylinder valve and remove the instrument from the
calibration cradle.
If a fault has occurred during the span adjustment:
The fault message »
respective sensor instead of the measured value.
In this case, repeat the adjustment.
Change the sensor if necessary.
7
Cleaning
The instrument does not need any special care.
Dirt and deposits can be removed from the instrument by washing it with
cold water. A sponge can be used for wiping if necessary.
!
CAUTION
Abrasive cleaning tools (brushes etc.), cleaning agents and cleaning solvents
can destroy the dust and water filters.
For information on suitable cleaning and disinfecting agents and their
specification, see document 9100081 at
www.draeger.com/IFU.
Carefully dry the instrument with a cloth.
« appears and »
Cleaning
« is displayed for the
en
47