Operation
!
WARNING
If an IR sensor is used in the Dräger X-am 5600, a adjustment of zero point and
sensitivity must be carried out after experiencing an impact load that results in a
non-zero display when exposed to fresh air.
!
WARNING
Incorrect measured values!
If water closes the gas inlets of the gas detector (e.g. by submerging the gas
detector under water or by heavy rain), this may result in incorrect measured
values.
Shake the gas detector with the display facing downwards in order to remove
the water.
The IP degrees of protection do not indicate that the gas detector will detect
a gas during or after exposure to this gas. In case of dust deposits and
contact with water due to submersion or a jet of water, check the adjustment
and functional integrity of the gas detector.
No additional accessories are required to ensure the IP degree of protection.
6.6

Identifying alarms

An alarm is displayed visually, audibly and through vibration in a specific pattern.
At low temperatures the legibility of the display can be improved by
switching on the backlight.
6.6.1
Concentration pre-alarm A1
Intermittent alarm:
Alternating display of » A1 « and measured value.
Not for O
!
2
The pre-alarm A1 is not latching and stops when the concentration has
dropped below the alarm threshold A1.
In case of A1, a single tone is audible and the alarm LED flashes.
In case of A2, a double tone is audible and the alarm LED flashes twice.
Acknowledge pre-alarm:
Press the [OK] key, only the audible alarm and the vibration alarm are
switched off.