4.3

Mounting the membrane body

– After calibration of the pH electrode rinse the pH
electrode with CO
– Fill the new membrane body half way with CO
electrolyte.
Make sure that all air bubbles are removed from the
membrane body. Air bubbles can be removed by care-
fully tapping on the membrane body.
– Slip the membrane body over the pH electrode
while holding the sensor in a vertical position.
The excess electrolyte will be displaced should
be wiped dry with a paper tissue.
h
Important: No electrolyte, sample media or
contamination may be present between the
membrane body and the cap sleeve. Please check
carefully.
– Carefully slip the cap sleeve over the fitted
membrane body, holding the sensor in a vertical
position and screw it tight. The cap sleeve must be
clean and tight.
© 03 / 2023 METTLER TOLEDO