Operation
5.1
Operation of the equipment
Once the sensor is properly calibrated (see § 6.4)
and installed, you should refer to the instruction
manual of the transmitter to learn how to operate the
complete system.
5.2
Measurement in power and failure
In case of power failure for more than 10-15 minu-
tes the sensor must be repolarized over a period of 6
hours (see Chapter 4.3).
In case of drop out for 5 to 15 minutes, the sensor
must be polarized for 45 minutes.
In case of a power drop-out less than 5 minutes, a
polarization time of 10 minutes is sufficient.
5.3
Errors and corrective actions
Make sure the sensor is mounted properly in a ver-
tical position (VP connector up), or at a 15° angle
(against the flow) to prevent the accumulation of air
bubbles on the sensor head.
5.4
Storage
The sensor and the membrane must be cleaned
before storing the sensor.
The sensor can be stored for several months, provi-
ded it is filled with O
cap is placed over the membrane. To avoid the
6-hour polarization requirement after storage, the
sensor can be stored connected to a polarization
module.