14
4.4
Mounting the sensor
Please read carefully the instruction manual of the
particular housing (InFit761-... or InTrac797) for
correct sensor installation.
The InPro 5000 (i) sensor can be mounted directly on
a top plate of a small bioreactor or in a pipe with an
inside Pg 13.5 thread and securely tightened via the
Pg 13.5 threaded sleeve.
h
Remark: Mounting upside-down or horizontaly is not
possible due to internal pH electrode. An angle of 15°
or bigger to horizontal is needed.
4.5
Sterilization
The installed sensor will be sterilized together with the
bioreactor. The InPro5000 is in-situ sterilizable or in an
autoclave. After cooling down but before inoculation a
process calibration step, as described in chapter 5, is
needed to achieve accurate measurements.
h
Remark: For best accuracy please wait one hour after
cooling down before process calibration step accor-
ding chapter 5.1.
InPro 5000 (i)
52 002 465 G
InPro 5000 (i) CO
© 03 / 2023 METTLER TOLEDO
Printed in Switzerland
Sensor 12 mm
2