OBJ_DOKU-189631-009.fm Seite 26 Mittwoch, 11. Mai 2022 3:22 15
Erbe recommends that the product should not be processed af-
ter use. Processing can affect the material properties and/or
function of the product such that use in accordance with the in-
tended use is no longer possible. Also, it cannot be ensured that
the product will be sterile after processing. Risks include:
• Components/connections that are damaged by high tem-
peratures (e.g. in the autoclave).
• Components (e.g., filter, extremely small lumen, etc.) that
may prevent internal cleaning/disinfection.
Erbe Elektromedizin expressly warns against modifying the
product. Any modification exempts Erbe Elektromedizin from
any and all liability.
4
Information on electromagnetic
compatibility (EMC) of the units
Notwithstanding the EMC specifications in the ERBEJET 2 and
VIO user manuals, HF emissions are classified as Class A
cording to CISPR 11 for common use of the units for the system
apply. All other EMC specifications in the User Manuals of the in-
dividual units also persist for common use of the units.
1.) The system is suited for use in environments other than domestic
areas and in ones directly connected to a puplic power supply sys-
tem that also supplies buildings being used for domestic purposes.
The customer or the user of the system should ensure that it is used
in such an environment.
1)