Herunterladen Inhalt Inhalt Diese Seite drucken

GF U3000 V2 Betriebsanleitung Seite 62

Ultraschall-durchflussmessgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 73
Troubleshooting
Unknown product
Illegal to edit or delete this information
ERR: unknown board type
Value out of bounds
System Error [nnnn]
62
Interpretation: The board count for product does not match the prod-
uct type specified.
Response: This is a serious error. Restart the
flowmeter. If the problem persists, contact your
distributor for further advice
Interpretation: This field cannot be modified or deleted. This usually
occurs when trying to edit or delete the Quickstart site.
Response: None required.
Interpretation: Flowmeter internal error. The controller has attempt-
ed to request a board that does not exist.
Response: Reset the flowmeter to be safe.
Record the conditions under which the error
occurred and report them to the distributor when
convenient.
Interpretation: The values entered were out of bounds for this vari-
able. This error is similar to the error "
Limits are xx.x [text] to yy.y [text]".
Response: Enter a valid value.
Interpretation: A serious internal error occurred. This indicates and
error condition that should not be
possible. It may or may not be fatal.
Response: Record the error number and conditions that lead to the
error. Ideally turn the flowmeter off then on. When convenient report
the error number and conditions to your convenient report the error
number and conditions to your distributor.