User Manual
4.4.2
Attaching the Guide Rail to the Pipe
Position the guide rail horizontally on the pipe at 45° with respect to the top of the pipe and secure it in position using the supplied
stainless steel banding, as shown in below.
UPSTREAM
NOTICE
In the following procedure the guide rail is installed with the rectangular opening facing towards the upstream end of the pipe
4.4.3
Fitting the Transducers
1. Tighten each transducer clamp clockwise until it is close to the top of the transducer
(figure below, left). This is necessary in order to prevent the acoustic couplant touching the pipe when the transducer is initially
inserted into the guide rail – as described below.
2. Using the supplied syringe applicator, apply a 3mm bead of acoustic couplant to the base of both transducers (figure below,
right).
3. Thread the downstream transducer cable (blue) through the right-hand end of the guide rail and up through the rectangular
opening at the top left-hand end of the guide rail, as shown in figure below.
Transducer clamp
Transducer
Cable connector
Couplant
DOWNSTREAM
Installation
21