Heat meter (HM versions only)

8
Heat meter (HM versions only)
NOTICE
This chapter applies to GF U3000 V2 HM models only.
Rtd Board
PT100 Sensors
Hot
46 °C
Energy
Cold
19 °C
Power
- >Calibrate Temperature Sensors. .
Exit
8.1

Calibrate temperature sensors

Connect the temperature sensors and check that the displayed values are reasonable.
54
DD-MM-YY HH:MM:SS
5.2107e+01 KJ
2.1784e+01 KW
1.
From the Main menu, use the Up and Down scroll keys to
select Setup Instrument. Press the ENTER Key. The Options
screen is displayed.
2.
Use the Up and Down scroll keys to select Heat-Meter..
Press the ENTER Key.
The Rtd-Board screen is displayed.
If sensors are connected, Hot and Cold temperatures will be
displayed. A display of "***" indicates no connection or a broken
sensor. The screen also displays the current total Energy and the last
measured instantaneous Power level.
1.
Clamp the sensors together and allow the
readings to stabilise.
2.
The sensors should be reading approximately the same
temperature. However, because of
small errors in the system, the values from each probe
may be fractionally different. If this is the case, the sensors
require calibration. To calculate power, it is the difference
in temperature that is important, rather than the absolute
temperature, although small differences in relative density
and specific heat capacity, which are a function of absolute
temperature are accounted for in the calculations.
Select Calibrate Temperature Sensors...
3.
4.
Enter the user PIN code (71360). The Calibrate Sensors
screen is displayed.
5.
For Use for Reference, select one of the
following:
Hot
Difference in readings between the two sensors is applied as
an offset to the Cold sensor.
Cold
Difference in readings between the two sensors is applied as
an offset to the Cold sensor.
Set Value
If you have an existing temperature measurement system and
trust its temperature reading. In this case, the Hot and Cold
sensors not only have to be clamped together, but they have
to be clamped to the point at which the existing equipment is
measuring temperature. Ensure that the temperatures have
stabilised.
None
Remove any offsets. If the temperature difference between the
two probes is more than 0.5 ⁰C, a power offset will be seen in
subsequent measurements.
6.
Select Calibrate.. The RTD Board screen is displayed. Check
that the temperature values now read the same value. A ✓
symbol is shown against the temperature reading that has
an offset associated with it and indicates that the probes
have already been calibrated.
User Manual