Outputs
How to calculate a suitable
1.Convert
Max. power
(eg. kW) to Max. energy per second
Example
Max. power = 5 kW
Max. energy per hour = 5kWh
Max. energy per second = Max. energy per hour (kWh) / 3600
Max. energy per second = 5000/3600 = 1.3888 kWh
Max. energy per second
Example
1.3888 kWh
2,
Calculate Max. pulses per second from
Example
Pulse width = 50 ms
Max. Pulses per second = 500 /
Max. Pulse per second = 500 /
6.2.2

Alarm output

An Alarm Output generates an alert when a predetermined value is exceeded or receded for Volume, Flow, Energy or Power, or
when Signal is lost or gained. When an alarm is activated a message is generated on the status line and the corresponding output
alarm symbol will flash.
50
Energy per pulse
value from
Max. pulses per second
Example
10
Maximum power
and
Pulse width
3.
Calculate
Example
Calculated:
Rounded up to next digit:
Rounded up to next decade:
pulse width (ms)
Pulse width (ms)
50ms
= 10
Select Output.. from the Options menu.
1.
2.
Use the UP/DOWN arrow keys to select Digital Device
1/2/3 Setup. Press the ENTER key. The Output 1/2/3 menu
is displayed.
3.
Use the UP/DOWN arrow keys to select Function. Press the
ENTER key.
4.
Use the UP/DOWN arrow keys to scroll through the output
types. Select Alarm Output.
5.
Use the UP/DOWN arrow keys to select Measurement
Source.
6.
Choose from Volume, Flow, Energy, Power or Signal.
7.
Use the UP/DOWN arrow keys to select Continue..
8.
According to your choice in step 6, complete the alarm
configuration as described in the following sections.
Volume Alarm
9.
Use the UP/DOWN arrow keys to select Direction. You may
select a Rising or Falling value (as volumes generally only
rise until reset, Rising is the usual choice).
10.
Use the UP/DOWN arrow keys to select Activation Level.
Set the volume limit for the alarm on this output.
If required, set a Deactivation Level although this has no
11.
effect until the volume totals are reset.
12.
Use the UP/DOWN arrow keys to select Save Setup and
Exit..
(Metric)
Energy per pulse
0.13888 kWh
0.2 kWh
1 kWh
User Manual