12.4 LoZ voltage mode
LoZ mode allows you to measure DC and AC voltages with a low impedance (ap-
prox. 400 kΩ). In this mode, the multimeter lowers the internal resistance to prevent
phantom voltage readings. As a result, the circuit is more heavily loaded than in the
standard measuring mode.
1. In order to use LoZ measurement mode, press the LoZ Low imp.400 kΩ button
during voltage measurement. The measured impedance is reduced for as long
as the button is pressed.
2. "LoZ" will appear on the display.
LoZ measurement mode may only be used up to a maximum voltage of
250 V. The duration of the LoZ measurement must be limited to a maxi-
mum of 3 s. This mode is not available in the mV measurement range.
After using LoZ mode, leave the multimeter for 1 minute before using it
again.