Herunterladen Inhalt Inhalt Diese Seite drucken

Mounting The Membrane Body - Mettler Toledo InPro 5000 Bedienungsanleitung

Co2 sensor
Vorschau ausblenden Andere Handbücher für InPro 5000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

52002465_F_IM_InPro5000i_edf_2022_09.qxp_52002465_F_IM_InPro5000_edf_2022
InPro 5000 (i) CO
4.3

Mounting the membrane body

– After calibration of the pH electrode rinse the pH
electrode with CO
– Fill the new membrane body half way with CO
electrolyte.
Make sure that all air bubbles are removed from
the membrane body. Air bubbles can be removed
by carefully tapping on the membrane body.
– Slip the membrane body over the pH electrode
while holding the sensor in a vertical position.
The excess electrolyte will be displaced should
be wiped dry with a paper tissue.
Important: No electrolyte, sample media or
contamination may be present between the
membrane body and the cap sleeve. Please check
carefully.
– Carefully slip the cap sleeve over the fitted
membrane body, holding the sensor in a vertical
position and screw it tight. The cap sleeve must be
clean and tight.
© 09 / 2022 METTLER TOLEDO
Printed in Switzerland
Sensor 12 mm
2
electrolyte solution.
2
1 / 2
1 / 2
CO
electrolyte
2
InPro 5000 (i)
52 002 465 F
13
2

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Inpro 5000i

Inhaltsverzeichnis