Troubleshooting
Error message
Err 4
Err 5
bAtt
46
Possible cause
1. Data transmission cable not
correctly inserted into the blood
pressure monitor
2. Pins in the plug of the data
transmission cable are
mechanically damaged.
3. Reading not transmitted
correctly
1. Power pack or battery voltage
too low
2. Power packs or batteries
defective
3. Battery contacts are corroded
Measures
1. Insert the cable correctly into the
blood pressure monitor (see
chapter 3.4 "Preparation for
measurement").
2. Check the plug to see whether
the pins on the inside are damaged.
If they are, call your specialist
supplier or IEM
3. Start the transmission again.
1. Replace the power packs or
batteries
2. The power pack or battery
voltage is correct but bAtt is shown
on the display during cuff inflation.
Replace the power packs
3. Clean the battery contacts with a
cotton cloth and a little alcohol.