The following should be displayed in this sequence:
Test
Battery status (volts)
Display Segment Test
Current 24-hour period
If an error occurs during the internal test, the blood pressure monitor will show E004 on the display and an audible
signal will sound. For safety reasons, the blood pressure monitor will be out of operation. Send the blood pressure
monitor immediately to your specialist supplier or directly to IEM GmbH for repair.

Clearing the memory

The memory must be empty before every measurement, i.e. there must be no blood pressure data from the
previous patient in the memory. If there are still values in there, these can be cleared using the delete function of
the evaluation software.
You can clear the memory manually by holding down the start button for more than 5 seconds. As it is held down,
all segments of the LCD will be displayed first of all, then 1 tone burst will be emitted, the number of saved
measurements will be displayed briefly and then "clr" will be displayed. If the event button is now pressed for more
than 2 seconds within 5 seconds, all measurements will be deleted.
Display
Comment
at least 2.6 V for NiMH power packs
2.85
at least 3.10 V for alkaline batteries
999:999
First the digits (999:999 to 000:000), then all other symbols of the
to
LCD appear in succession.
000:000
Check whether all segments are correctly and fully displayed (the
complete program code is checked for correctness in the
background).
21:45
Between 00:00 and 23:59