5.4 Mobil-O-Graph
error description
®
Error message
Err 1
Err 2
Err 3
Possible cause
1. The patient has severe
arrhythmia
2. Arm moved during
measurement
3. Insufficient valid pulse rate
detected
1. Arm moved during
measurement
2. Cuff does not fit the arm snugly
1. Blood pressure is beyond the
measurement range
2. Considerable arm movement
3. Problems with the pneumatics
Troubleshooting
Measures
1. Blood pressure monitor not
applicable
2. Keep the arm still during
measurement
3. Place the cuff on the arm again
1. Keep the arm still during
measurement
2. Check the seating of the cuff and
that of the device
1. Permanent notifications render
the blood pressure monitor
unsuitable for the patient.
2. Keep the arm still during
measurement.
3. If this error occurs continuously,
send the device to your specialist
supplier or directly to IEM GmbH for
checking.