Product description
Interfaces:
Expected operational life of the
device
Expected operational life of the
cuff
Environmental conditions:
ATTENTION
Extreme temperatures, humidity or air pressure can influence the measuring accuracy. Please
observe the operating conditions.
Extreme temperatures, humidity or altitude can affect the performance of the blood pressure
monitor. Do not store the device near a fireplace or radiant heater and do not expose it to extreme
sunlight. Do not place the device next to a nebulizer or a steam boiler, as condensation may
damage the device.
The blood pressure monitor needs about 25 minutes to go from the minimum storage temperature
of -25 °C to the operating temperature of 5 °C at an ambient temperature of 20 °C.
It takes approximately 25 minutes for the blood pressure monitor to go from the maximum storage
temperature of 70 °C to the operating temperature of 40 °C at an ambient temperature of 20 °C.
36
IR-MED serial or USB (IEM specific)
PC combination cable USB or serial
Bluetooth
5 years
6 months