Herunterladen Inhalt Inhalt Diese Seite drucken

VATECH A9 Handbuch Seite 40

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 157
3. Acquiring PANO Images
Laser Beam Aligning
1.
Align the Vertical Beam with the center of the face (Mid-sagittal Line). (It is to
prevent the horizontal expansion of the image)
2.
Align the Horizontal Beam in a straight line to the Frankfurt Line on the patient's
face. Use the Horizontal Beam button on the control panel to position it. Make
sure that the Horizontal Beam is aligned to the patient's face horizontally.
Finishing Patient Positioning
1.
Click the READY button. X-ray exposure has not started yet.
2.
Now go to 3.3 X-ray Exposure to start the exposure.
34
Be careful not to shine the laser beam directly into the
person's eyes. Doing so may result in vision loss.
If the laser beam is not correctly positioned, there may be
distortion, causing the image to be enlarged or reduced, or
ghost shadows may occur and lower the image quality. Be
sure to align the Laser Beam properly.
vatech A9 (PHT-30CSS) User Manual

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Pht-30css

Inhaltsverzeichnis