8.
Guide the patient to sit in the center of the equipment and maintain the position
described below.
•
Hold the handle firmly.
•
Make sure the patient's chest is in contact with the device.
•
Keep both shoulders parallel.
•
Straighten the patient's cervical spine and sit still.
9.
Guide the patient to press the base of the nose (acanthion point) against the
Chinrest and tilt the head forward about 5˚. At this point, make sure the patient's
jaw does not touch the equipment.
10. Let the patient maintain the posture as follows:
•
Close the mouth.
•
Place the tongue to the roof of the mouth.
•
Close the eyes.
vatech A9 (PHT-30CSS) User Manual
▪
If the jaw touches the equipment, it is difficult to maintain
the proper position to get good images.
▪
Be careful; the patient does not touch the equipment with
his/her jaw.
▪
As shown in the picture, the support unit of the integrated
Chinrest should touch the patient's acanthion point.
▪
Ask the patient to maintain his/her position until the
operation is completed.
3. Acquiring PANO Images
33