3.2.2
SPECIAL Examination Mode (TMJ/Sinus)
[TMJ OPEN Mode (LAT)]
The TMJ Close image can be acquired after the TMJ Open image is acquired.
Patient Positioning
1.
Remove the Normal Chinrest and insert the Special Chinrest into the
equipment.
2.
Insert the Special Bite A into the Special Chinrest.
3.
Guide the patient to a chair (sold separately) in the center of the equipment.
4.
Guide the patient to sit in the center of the equipment
vatech A9 (PHT-30CSS) User Manual
Steps for TMJ Mode
Patient positioning for TMJ Open → Laser Beam Aligning
→ X-ray Exposure → Patient positioning for TMJ Close →
Laser Beam Aligning → X-ray Exposure
Special Bite A
Special Chinrest
Clean the Chinrest and the Bite with non-alcohol based, non-
corrosive cleaning solution and wipe with a dry towel before
the next patient.
If a chair with a backrest is used, make sure that the
patient's back is away from the backrest of the chair for
proper positioning.