3. Acquiring PANO Images
[TMJ CLOSE Mode (PA) and Sinus Mode (LAT)]
Patient Positioning
1.
Remove the Normal Chinrest and insert the Special Chinrest into the
equipment.
2.
Insert the Special Bite B into the Special Chinrest.
3.
Guide the patient to a chair (sold separately) in the center of the equipment.
4.
Guide the patient to sit in the center of the equipment
5.
Position the patient's back and cervical spine as straight as possible.
6.
Adjust the instrument to the patient's sitting height using the Column UP/DOWN
on the Handle frame or the optional UP/DOWN switch while the patient holds the
handle.
7.
Close contact patient's chest to the handle frame.
32
Special Bite B
Special Chinrest
Clean the Chinrest and the Bite with non-alcohol based, non-
corrosive cleaning solution and wipe with a dry towel before
the next patient.
If a chair with a backrest is used, make sure that the
patient's back is away from the backrest of the chair for
proper positioning.