Callisto™ – Instructions for Use - EN
Note that if too many curves are f rozen on the screen not all will be
saved in Noah due to imposed limitations.
List of Protocols allows you to select a test protocol (def ault or user
def ined) to use in the current test session.
Temporary Setup button allows f or making temporary changes to the
selected test protocol. The changes will be valid f or the current session
only. Af ter making the changes and returning to the main screen, the
name of the test protocol will be f ollowed by an asterisk (*).
List of Historical Sessions accesses historical sessions f or
comparison or printing purposes.
Toggle between Lock and Unlock the Selected Session f reezes the
current or historical session on the screen f or comparison to other
sessions.
Go to Current Session button brings you back to current session.
Toggle between Coupler and Ear button allows you to toggle between
real-ear and coupler/test box mode.
NOTE: This icon only becomes active if a RECD (measured or
predicted) is available.
Report Editor button opens a separate window f or adding notes to the
current session. Af ter saving the session, changes can only be made
within the same day until the date changes (at midnight). Note: these
timef rames are limited by HIMSA and the Noah sof tware, and not by
Interacoustics.
Single Frequency button is a test that lets the f itter
play a single f requency warble tone. Once clicked, the
exact f requency, input and output can be seen on the
graph. The f requency can be adjusted up and down by
using the Right and Lef t arrows on the keyboard. Click
on the button to turn it on and click on it again to turn it
of f .
UCL (Uncomfortable Levels) Adjustment
To limit the system's signal intensity while measuring the MPO during a
Real-Ear situation, the UCL button can be activated. Once activated, a
red line will appear on the graph and the system will stop measuring if
this UCL level is reached. This red line can be adjusted with the slider.
NOTE: UCL thresholds must be entered on the audiogram f or the red
line to appear when the UCL button is active. To deactivate this f eature,
press on the UCL button again.