HF requires an additional feature for the AC440. If not purchased, the button is greyed out.
2
MF requires additional feature for the AC440. If not purchased, the button is greyed out.
Callisto™ – Instructions for Use - EN
Fold an area so that it only shows the label or the buttons of that area.
Unfold an area so that all buttons and labels are visible.
Show/hide areas can be f ound by right mouse clicking on one of the
areas. The visibility of the dif ferent areas as well as the space that they
take on the screen is locally saved to the examiner.
List of Defined Protocols allows f or selecting a test protocol f or the
current test session. Right mouse click on a protocol allows the current
examiner to set or deselect a def ault startup protocol. Please ref er to the
"Additional Inf ormation" document f or more inf ormation about protocols
and protocol setup.
Temporary Setup allows f or making temporary changes to the selected
protocol. The changes will be valid f or the current session only. Af ter
making the changes and returning to the main screen, the name of the
protocol will be f ollowed by an asterisk (*).
List of historical sessions accesses historical sessions f or comparison
purposes. The audiogram of the selected session, indicated by the
orange background, is shown in colours as def ined by the used symbol
set. All other audiograms that are selected by check marks show on
screen in the colours as indicated by the text colour of the date and time
stamp. Note that this listing can be resized by dragging the double lines
up or down.
Go to Current Session brings you back to the current session.
High Frequencies
1
show f requencies on the audiogram (up to 16 kHz f or
the Callisto™). However, you will only be able to test in the f requency
range the high f requency headset is calibrated f or.
High Frequency Zoom activates high f requency testing and zooms in on
the high f requency range. Please ref er to the Callisto™ "Additional
Inf ormation" document f or more inf ormation on HF Audiometry
Toggle between Single and Dual Audiogram button toggles between
viewing the audiogram as one combined graph or two separate graphs.
Multi Frequency
2
activates Multi Frequency testing. The f requency steps
can be adjusted in the AC440 setup.
Synchronize channels locks the two channels together. This f unction
may be used to perf orm synchronous masking.