Herunterladen Inhalt Inhalt Diese Seite drucken

Electro-Magnetic Compatibility Information; Electro-Magnetic Emissions; Elector-Magnetic Immunity - KULZER cara i600 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 137

6 Electro-Magnetic Compatibility Information

6.1 Electro-Magnetic Emissions

This i600 is intended for use in the electromagnetic environment as
specified below. The customer or the user of the i600 should ensure
that it is used in such an environment.
Emission Test
Compliance
RF Emissions
Group 1
CISPR 11
RF Emissions
Class A
CISPR 11
Harmonicemissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Complies
Flicker emissions
Warning : This i600 is intended for use by healthcare professionals
only. This equipment/ system may cause radio interference or may
disrupt the operation of nearby equipment. It may be necessary to
take mitigation measures, such as re-orienting or relocating the i600
or shielding the location.
18
User Manual
Electromagnetic
Environment - Guidance
The i600 uses RF energy only for
its internal functions. Therefore,
its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
The EUT is suitable for use in all
establishments. This includes
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
6.2 Electro-Magnetic Immunity
G uidance 1
... ...
... T his i600 is intended for use in the electromagnetic environment
as specified below. The customer or the user of the i600 system
should ensure that it is used in such an environment.
Immunity
IEC 60601
Test
test level
Electrostatic
± 8 kV contact ±
± 8 kV contact ±
discharge
2 kV, ± 4 kV, ± 8
2 kV, ± 4 kV, ± 8
(ESD) IEC
kV, ± 15 kV air
kV, ± 15 kV air
61000-4-2
±2 kV (for power
±2 kV (for power
Electrical fast
supply lines) ±
supply lines) ±
transient/burst
1 kV (for input/
1 kV (for input/
IEC 61000-4-4
output lines)
output lines)
±0.5 kV, ±1 kV
±0.5 kV, ±1 kV
differential mode
differential mode
Surge IEC
±0.5 kV, ±1 kV,
±0.5 kV, ±1 kV,
61000-4-5
±2 kV common
±2 kV common
mode
mode
Electromagnetic
Compliance
environment ‐
level
guidance
Floors should be
made of wood,
concrete, or ceramic
tiles. If floors are
covered with a
synthetic material,
a relative humidity
of at least 30% is
recommended.
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Mains power quality
should be that of a
typical commercial
or hospital
environment.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis