6 Electro-Magnetic Compatibility Information

6.1 Electro-Magnetic Emissions

This i700 is intended for use in the electromagnetic environment as
specified below. The customer or the user of the i700 should ensure
that it is used in such an environment.
Emission Test
Compliance
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonicemissions
IEC 61000-3-2
Voltage fluctuations/
Flicker emissions
Warning : This i700 is intended for use by healthcare professionals
only. This equipment/ system may cause radio interference or may
disrupt the operation of nearby equipment. It may be necessary to
take mitigation measures, such as re-orienting or relocating the i700
or shielding the location.
18
User Manual
Electromagnetic
Environment - Guidance
The i700 uses RF energy only for
its internal functions. Therefore,
its RF emissions are very low
Group 1
and are not likely to cause any
interference in nearby electronic
equipment.
Class A
The EUT is suitable for use in all
establishments. This includes
domestic establishments and
Class A
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Complies
6.2 Electro-Magnetic Immunity
G uidance 1
... ...
... T his i700 is intended for use in the electromagnetic environment
as specified below. The customer or the user of the i700 system
should ensure that it is used in such an environment.
Immunity
IEC 60601
Test
test level
Electrostatic
± 8 kV contact ±
discharge
2 kV, ± 4 kV, ± 8
(ESD) IEC
kV, ± 15 kV air
61000-4-2
±2 kV (for power
Electrical fast
supply lines) ±
transient/burst
1 kV (for input/
IEC 61000-4-4
output lines)
±0.5 kV, ±1 kV
differential mode
Surge IEC
±0.5 kV, ±1 kV,
61000-4-5
±2 kV common
mode
Electromagnetic
Compliance
environment ‐
level
guidance
Floors should be
made of wood,
concrete, or ceramic
± 8 kV contact ±
tiles. If floors are
2 kV, ± 4 kV, ± 8
covered with a
kV, ± 15 kV air
synthetic material,
a relative humidity
of at least 30% is
recommended.
Mains power quality
±2 kV (for power
should be that of a
supply lines) ±
typical commercial
1 kV (for input/
or hospital
output lines)
environment.
±0.5 kV, ±1 kV
Mains power quality
differential mode
should be that of a
±0.5 kV, ±1 kV,
typical commercial
±2 kV common
or hospital
mode
environment.