Screwing with GB896R/GB897R
■
10 second mechanical application, 10 second manual application,
30 second pause
■
30 repetitions
■
30 min cooling time
■
Max. Temperature 48 °C
12.4 Environmental conditions
Operation
Temperature
10 °C to 27 °C
Relative humidity
30 % to 75 %
Atmospheric pres-
700 hPa to 1 060 hPa
sure
13. Disposal
Note
The user institution is obliged to process the product before its disposal, see
Validated reprocessing procedure.
Adhere to national regulations when disposing of or recy-
cling the product, its components and its packaging!
The recycling pass can be downloaded from the Extranet
as a PDF document under the respective article number.
(The recycling pass includes disassembling instructions for
the product, as well as information for proper disposal of
components harmful to the environment.)
Products carrying this symbol are subject to separate col-
lection of electrical and electronic devices. Within the
European Union, disposal is taken care of by the manufac-
turer as a free-of-charge service.
►
Detailed information concerning the disposal of the product is avail-
able through your national B. Braun/Aesculap agency, see Technical
Service.
Storage and transport
-10 °C to 50 °C
10 % to 90 %
500 hPa to 1 060 hPa
en
15