Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Technical Information
10 Disable the "TIP-SEAL SERV" warning message by serial interface. To disable the Tip-
seal Service message you have to reset the Tip-seal hour-meter (Win 358). Write a "0"
on the window 358. The message will be showed again after 5000 h of operation. To
check the Tip-seal life you have just to read the Win 358.

To test the Pumping System

1
Run the TPS-compact for about 5 seconds. Verify that the front fan is running.
If you hear loud noises from the Scroll pump this indicates that the tip seal or
main O-ring are possibly out of place.
2
Disassemble and repair as necessary.
3
The pump is now ready to return to service.
Newly installed tip seals may require several hours of run time to seat properly and
NOTE
enable the pump to meet speed and base pressure specifications.
230
Tip Seal Life Reset using A-plus
Agilent TPS-compact User Manual