Electrical connection
8.1
Connection in a six-wire configuration
Fig. 8.1
Pin assignment in a six-wire circuit
With this cable assignment, the output voltage at the measuring amplifier is
positive in the tensile direction when the transducer is loaded.
The cable shield is connected with the transducer housing. Only use plugs that
meet EMC guidelines. The shielding must be connected extensively. With oth
er
connection techniques, an EMC-proof shield should be applied in the stranded
wire area and this shielding should also be connected extensively (also see
HBM Greenline Information, brochure i1577).
28
Device plug pin assignment
B
D
C
A
F
E
A
F
B
E
C
D
A01385_06_YIC_06 HBM: public
Measurement
signal (+)
Excitation voltage (-)
Measurement signal (-)
Excitation voltage (+)
Sense lead (+)
Sense lead (-)
Cable shield, connected
to the housing
Cable assignment
wh (white)
bk (black)
rd (red)
bu (blue)
gn (green)
gy (gray)
Shield
U10M/U10S