Mechanical installation
M16
22
16,5 f7
M16
A
You do not need to pay attention to this information if measurement takes
place at uniform temperatures or taring is possible before changes in tempera
ture. Likewise, large nominal (rated) forces above 25 kN are not affected.
The screwed-in force application parts must be sufficiently pre-stressed to
meet the stated specifications and the full endurance strength of the threads.
This is implemented with a locknut screwed onto the customer-side structural
element. We recommend using one of the methods described below.
1. Pre-stressing with defined tightening torque
Force transducers with a nominal force up to and including 25 kN can be
mounted using this method.
S Screw the force application part into the transducer or adapter as far as the
retainer.
S Unscrew force application part two revolutions.
S Pre-stress force application part with defined tightening torque.
20
A01385_06_YIC_06 HBM: public
min. 22
16,5 f7
0,02
M16
A
2:1
A
0,03 A
U10M/U10S