12.3 Guidelines and Manufacturer's Statement – Electromagnetic
Emissions
The Aquilex® Fluid Control System is intended for use in an environment as de-
scribed below. The user/operator of the Aquilex® Fluid Control System should
make sure the device is operated within such an environment.
Emitted interference measurements
HF emission according to CISPR 11
HF emission according to CISPR 11
Emission of harmonic oscillations accord-
ing to IEC 61000-3-2
Emission of voltage fluctuations / flickers
according to IEC 61000-3-3
12.4 Guidelines and Manufacturer's Statement/Electromagnetic
Interference Immunity
The Aquilex® Fluid Control System is intended for use in an electromagnetic envir-
onment as described below. The user/operator of the Aquilex® Fluid Control Sys-
tem must make sure the device is operated within such an environment.
Electromagnetic interference
immunity tests
Discharge of static electricity
(ESD)
according
to
IEC 61000-4-2
Electrical
fast
transients/
bursts
according
to
IEC 61000-4-4
Surges
according
to
IEC 61000-4-5
Blackouts,
brownouts,
and
fluctuations of the power sup-
ply
according
to
IEC 61000-4-11
Supply frequency magnetic
field (50/60  Hz) according to
IEC 61000-4-8
Conducted
HF
interference
quantities
according
to
IEC 61000-4-6
Radiated
HF
interference
quantities
according
to
IEC 61000-4-3
Compliance
Group 1
Class B
Class A
In compliance
Test level
± 6 kV contact discharge
± 8 kV air discharge
± 2 kV for power lines
±  1  kV for input and output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
< 5 % U
* (> 95 % dip in the U
T
T
for ½ cycle
< 40 % U
(> 60 % dip in the U
T
T
for 5 cycles
< 70 % U
(> 30 % dip in the U
T
T
for 25 cycles
< 5 % U
(> 95 % dip in the U
T
T
for 5 cycles
3 A/m
3 V
150 kHz to 80 MHz
eff
3 V/m 80 MHz to 2.5 GHz
Electromagnetic environment guidelines
The Aquilex® Fluid Control System uses
HF energy solely for its internal func-
tions. Therefore, its HF emission is very
low and it is unlikely that devices in close
proximity will experience interference.
The Aquilex® Fluid Control System is suit-
able for use in all facilities including
those in residential areas and those dir-
ectly connected to a public utility net-
work supplying buildings used for resid-
ential purposes as well.
Compliance
In compliance
In compliance
In compliance
)
)
)
In compliance
)
In compliance
In compliance
In compliance
Electromagnetic Compatibility
Electromagnetic environment
guidelines
Floors should be made from
wood or concrete or covered
with ceramic tiles. If the floor
covering consists of synthetic
material, the relative humidity
should be at least 30 %.
The quality of the supply
voltage should be the same as
the voltage of a typical busi-
ness or hospital environment.
The quality of the supply
voltage should be the same as
the voltage of a typical busi-
ness or hospital environment.
The quality of the supply
voltage should be the same as
the voltage of a typical busi-
ness or hospital environment.
If the user/operator of the sys-
tem requires the continuation
of functionality after power
interruptions/disruptions, it is
recommended to supply the
device with power from an un-
interruptible power supply.
Magnetic fields of the mains
power frequency should com-
ply with the typical values of
business and hospital environ-
ments.
Portable and mobile wireless
devices should not be used in
closer proximity to the device
Aquilex® Fluid Control System
(including cables/lines) than
the recommended safety dis-
57 / 412
en

Werbung