Herunterladen Diese Seite drucken

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 70

Technical Data

en
Model or type designation
Manufacturer information
Software version
Mains voltage range [V]
Supply frequency range [Hz]
Fuse designation
Internal voltage supply
Max. allowed load of the additional socket / more than one
sockets [A or VA]
Power consumption
Upper voltage range
Lower voltage range
Protection class (I, II, III)
Application part type (B, BF, CF)
Defibrillator protected (yes/no)
Protection type (IP code)
Classification (I, IIa, IIb, III) acc. to Appendix  IX of European
MDD
Conformity with the following standards
Operating conditions
Use possible with flammable anesthetic gases
Storage conditions [°C] [°F], [%], [kPa]
Transport conditions [°C] [°F], [%], [kPa]
Max. sound level
Maximum flow rate [l/min]
Maximum load
Maximum negative suction pressure
Adjustable values
52 / 412
10
Normal operation
Peak
Normal operation
Peak
Pressure range [mmHg]
Deficit limit [l]
Suction pressure [ kPa]
Technical Data
Aquilex Fluid Control System ( REF: AQL-100)
consisting of:
Irrigation Pump Unit (REF: AQL-100P)
Fluid Monitoring Unit (REF: AQL-100CS)
W.O.M. WORLD OF MEDICINE GmbH
Salzufer 8, 10587 Berlin
See user manual/operator instructions to determine software
version
100 - 240 V~
50/60 Hz
2x T 3.15 AH, 250 V, UL- recognized
No
1.6 A
Current [A]
0.19 A
0.69 A
0.52 A
1.70 A
I
BF (with Aquilex Fluid Control System inflow tube set)
No
IP41 (AQL-100P), IP21 (AQL-100CS)
IIb
EN 60601-1:2006 / IEC 60601-1:2005
EN 60601-1-2:2007 / IEC 60601-1-2:2007
10 to 40 °C / 50 to 104 °F
30 to 75 % rel. humidity
70 to 106 kPa air pressure
3000 m max. altitude above sea level for device use
This system is not designed for use with flammable anesthetic
agents (Class AP) or flammable anesthetic agents with oxid-
ants (Class APG).
5 to 40 °C/41 to 140 °F
5 to 85 % rel. humidity
70 to 106 kPa air pressure
-20 to 70 °C/-4 to 158 °F
5 to 90 % rel. humidity at 30 °C/86 °F
70 to 106 kPa air pressure
< 80 dB(A)
0.8 l/min ± 10 %
65 lbs/30 kg
-67 kPa
40 - 150 mmHg
0.6 – 2.5 l
0, LOW: -30 kPa (fixed value), HIGH: -40 to -67 kPa
Power consumption
Voltage [V]
[VA/W]
240 V
45 VA
240 V
165 VA
100 V
52 VA
100 V
170 VA
1

Werbung

loading