► Report each serious incident in connection with the product, in
particular a worsening of the state of health, to the manufacturer
and to the relevant authority in your country.
► Please keep this document for your records.
1.1 Construction and Function
The 1D10 and 1D11 prosthetic feet are suitable for use in both modu­
lar and exoskeletal prostheses. The functional properties are achieved
through the combination of a contoured core and functional foam.
The 1D10 prosthetic foot is available either without a foot adapter or
with a preinstalled titanium foot adapter. An ankle block can only be
installed without the preinstalled foot adapter.

1.2 Combination possibilities

This prosthetic component is compatible with Ottobock's system of
modular connectors. Functionality with components of other manufac­
turers that have compatible modular connectors has not been tested.
Allowable ankle blocks and screw connections
Reference number
(prosthetic foot)
1D10 (without adapter)
1D11
Allowable foot adapters
Reference number
(prosthetic foot)
1D10 (without adapter)
1D11
Foot size
Reference number
[cm]
(ankle block and screw
connection)
All sizes
2K34* and 2Z22=M10
22 to 25
2K34* and 2Z22=M8x90
26 to 28
2K34* and 2Z22=M10
Foot size
Reference number
[cm]
(foot adapter)
All sizes
2R8=M10, 2R31=M10,
2R54=M10
22 to 25
2R8=M8, 2R31=M8,
2R54=M8
26 to 28
2R8=M10, 2R31=M10,
2R54=M10

2 Intended use

2.1 Indications for use
The product is intended exclusively for lower limb exoprosthetic fit­
tings.
2.2 Area of application
Our components perform optimally when paired with appropriate
components based upon weight and mobility grades identifiable by
our MOBIS classification information and which have appropriate
modular connectors.
The product is recommended for mobility grade 1 (indoor
walker) and mobility grade 2 (restricted outdoor walker).
kg
The maximum approved body weight is specified in the technical
data (see Page 13).

2.3 Environmental conditions

Allowable environmental conditions
Temperature range: –10 °C to +45 °C
Moisture: relative humidity: 20% to 90%, non-condensing
Unacceptable environmental conditions
Chemicals/liquids: fresh water, salt water, perspiration, urine,
acids, soapsuds, chlorine water
Solids: dust, sand, highly hygroscopic particles (e. g. talcum)
Other environmental conditions
UV light
Hydrolysis
Storage and transport
Temperature range –20 °C to +60 °C (–4 °F to +140 °F), relative
humidity 20 % to 90 %, no mechanical vibrations or impacts
Not resistant
Not resistant