Herunterladen Inhalt Inhalt Diese Seite drucken

Construction And Function; Combination Possibilities; Intended Use; Environmental Conditions - Ottobock 1S49 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für 1S49:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1

1.1 Construction and Function

The 1S49, 1S66, 1S67 and 1S90 SACH (Solid Ankle Cushion Heel)
prosthetic feet are suitable for use in both modular prostheses and
exoskeletal prostheses. The functional properties are achieved
through the combination of a contoured core and functional foam.

1.2 Combination possibilities

This prosthetic component is compatible with Ottobock's system of
modular connectors. Functionality with components of other manufac­
turers that have compatible modular connectors has not been tested.
Allowable ankle blocks and screw connections
Foot size [cm]
Reference number (ankle block and screw
24 to 30
Permissible foot adapters
Prosthetic
Size [cm]
foot
1S67
22 to 25
26
26
1S49, 1S66,
21 to 30
1S90
26 to 30
10
connection)
2K34=30 and 2Z22=M10
Body weight
Permissible foot
[kg]
adapter
≤100
2R8=M8, 2R31=M8,
2R54=M8
2R8=M10,
2R31=M10,
2R54=M10
101 to 125
2R8=M10,
2R31=M10
≤100
2R8=M10,
2R31=M10,
2R54=M10
101 to 125
2R8=M10,
2R31=M10

2 Intended use

2.1 Indications for use
The product is intended exclusively for lower limb exoprosthetic fit­
tings.
2.2 Area of application
Our components perform optimally when paired with appropriate
components based upon weight and mobility grades identifiable by
our MOBIS classification information and which have appropriate
modular connectors.
The product is recommended for mobility grade 1 (indoor
walker) and mobility grade 2 (restricted outdoor walker).
kg
The maximum approved body weight is specified in the technical
data (see Page 14).

2.3 Environmental conditions

Allowable environmental conditions
Temperature range for use: -10 °C to +60°C
Allowable relative humidity 0 % to 90 %, non-condensing
Unallowable environmental conditions
Mechanical vibrations or impacts
Perspiration, urine, fresh water, salt water, acids
Dust, sand, highly hygroscopic particles (e.g. talcum)
Other environmental conditions
UV light
Hydrolysis

2.4 Service life

This product has been tested by the manufacturer with 2 million load
cycles. Depending on the patient's activity level, this corresponds to a
service life of 2 to 3 years.
Not resistant
Not resistant

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

1s661s671s90 sach

Inhaltsverzeichnis