Before entering the contaminated areas, remove the Seal Check Cap by gently twisting the handle so that one side
of the cap comes loose from the exhalation valve. Be careful not to dislodge the exhalation valve cover. See
Figure 21
.
Be sure to remove the Seal Check Cap before entering the contaminated area. The Cap blocks
the exhalation valve, making it difficult for exhaled air to be expelled from the mask. Failure to
remove the cap will lead to build-up of Carbon Dioxide in the mask, and may result in

11. Using CleanSpace HALO With Full Face Mask

Complete the following five (5) steps each time you use your respirator with the Full Face Mask.
STEP 1 - INSPECT
Before entry into a contaminated area, the following inspections must be performed.
Visually check the all parts including the Respirator, mask, harness and filter. If parts are missing or damaged
replace only with approved parts before proceeding.
Check the RESPIRATOR for cracks, holes or other damage or missing parts. Do not use the device if there is
any damage or misuse. Check both bellows for splits or holes. Check that the bellows have not become
distorted so as to partially or fully close the air path to the mask. Check and fit the neck support (if required).
Check the battery is fully charged by pressing the Power button. All three (3) battery Green LEDs must light
up. See Section 14.
Check the FILTER carefully. The foam seal must be clean and free from damage of any kind. Examine the
visible internal surfaces for any sign that dust has penetrated the media. If dust is found, do not use the filter.
The body of the filter must be free of cracks or signs of damage. If any sign of impact or scratching is found,
discard the filter. Fit the filter to the Respirator (see Changing the Filter, below)
FILTERS ARE DAMAGED BY THE USE OF COMPRESSED AIR OR BY BRUSHING.
Misuse of CleanSpace Respirator by cleaning the filter may result in overexposure to
Check the FULL FCE MASK to ensure that there are no cracks, tears or dirt; Check the mask is not distorted.
Check the mask exhalation valve for damage or dirt build up. If it is dirty, remove the exhalation valve cover.
Remove any dirt, hairs or anything that could affect the seal of the valve against its seat. Check that the valve
seat is clean. Reinstall the valve cover. If the valve is damaged, replace the mask.
Check the HEAD HARNESS is intact and has good elasticity and the fastening clips can be tightened and
loosened. The harness must be adjusted to support the mask to seal comfortably to the face. DO NOT
overtighten. Finally check the airclips can be clicked/unclicked into the Respirator.
STEP 2 – CALIBRATE
headache or dizziness. Never leave the Cap in place for more than 2 minutes!
DO NOT USE COMPRESSED AIR OR A BRUSH TO CLEAN THE FILTER!
contaminants and lead to sickness or death
Figure 21 - Removing Seal Check Cap
17
.