For the battery to be 100% charged plug in the charger until all three (3) Green LEDs are on and solid (no
Green LEDs
*
Operating time is affected by mask seal, work rate and other factors. The operating times quoted above
are average durations at moderate work rates at sea level. Actual operating times may vary widely from
quoted average durations.

CHARGING THE BATTERY

The battery must only be charged with a CleanSpace Battery
Charger.
To charge the internal battery, turn the respirator upside
down.
Locate the black charging port. See Figure 24 - Charging
Port
Insert the charger connector into the charging port.
Ensure the charger is plugged into a power source.
The Battery Indicator Green LEDS on the keypad should
be lit up and flashing as the Respirator charges.
The Respirator can be charged at temperatures
between 0°C to 35°C. The battery will not charge
outside this temperature range.
Check that at least one of the Battery Indicator Green LEDS on the keypad is on and flashing. If none of
the Green LEDs are flashing, the unit is not charging. Check the charger is connected into the Respirator
and the power source. If the respirator is still not charging, contact CleanSpace Customer Service
sales@cleanspacetechnology.com or visit our website:www.cleanspacehealth.com
Charging is complete when the respirator shows three (3) green solid lights. If the Respirator is
unplugged when the third Green LED is flashing, charging is 95% complete.
When the respirator is fully charged, disconnect the charger cable from the Respirator charging port.
IMPORTANT: The Respirator should NOT be worn when on charge.
Do not charge the battery in electro-magnetic environments as it may damage the electronics.

LOW BATTERY ALARM

CleanSpace respirators have a Low Battery Alarm that is both audible (3 beeps, repeated every second) and visual
(Green LEDS) that triggers when the battery has low battery charge. On moderate work rates, this may be 5 – 15
minutes run time (depending on filter loading). Unlike the filter alarm, the Low Battery Alarm cannot be muted. The
Alarm will turn off when the respirator is charging or by switching the Respirator off. When the battery charge is
extremely low, the motor and airflow will stop.
If the battery alarm sounds (3 beeps, repeated every second), leave the contaminated area
immediately and re-charge the battery. Operating the respirator after the low battery alarm
has sounded can cause the flow to fall below the manufacturer's minimum designed flow,
which may result in mask leak and exposure to contaminants that may lead to sickness or
flashing).
With the unit NOT on charge
Charge
85 - 100%
15 – 85%
5 – 15%
death
25
Approximate Run Time*
Recharge! (20min to 1 hour)
Figure 24 - Charging Port
.
4 – 8 hours
1 – 4 hours

Werbung