This respirator contains a system for synchronising with your breathing and regulating mask pressure. The system
should be re-calibrated if it experiences a significant change in temperature of more than 20°C. It is best
practice to re-calibrate if the unit has been in storage when the storage temperature is unknown.
Steps to re-calibrate:
6.
Remove the filter from the respirator.
With CleanSpace Respirator in Standby Mode (ie one or more green LEDs lit), press the "Power" and "Test
7.
Flow" buttons on the keypad at the same time.
8.
The red and blue LEDs will both light, and the green battery LEDs will cycle.
Hold the respirator still until the red and blue LEDs turn off (10 – 15 seconds). After this, calibration is complete.
9.
10. At the end of the calibration process the green LEDs return to indicating battery charge status.
STEP 3 – TEST FLOW RATE
This test checks that the Respirator is able to deliver the Manufacturer's Minimum Design Flow of 120 litres/minute.
5.
Remove the mask from the respirator. Place the respirator flat on a table or other support.
Press and release the button marked "Flow Test"
6.
7.
The respirator will automatically run the Flow Test. Note: During the test, the motor will run fast and air will
blow from the left hand bellows.
8.
After 2 seconds the respirator reports the Test result using the LEDs on the keypad. Use the table below to
interpret the LEDs.
LIGHTS
Meaning
PASS (Excellent: flow >180 l/min)**
3 LEDs:
2 LEDs:
PASS (Good)**
1 LED:
PASS (Acceptable)
FAIL (Flow <120 l/min)
ALL LEDs
FLASH
Do not use the respirator until a new filter has been fitted and / or battery fully charged and the test has
been repeated with a PASS result. Reset the respirator by pressing the Power button. Fully charge the
battery and / or replace the filter. Repeat the flow test. If the respirator fails the test filter is new and battery fully
charged but, contact CleanSpace Technology and do not use until it has been evaluated
** WARNING: The Test Flow Rate is not a battery charge test. Three LEDs means that, as of the moment it is tested, the unit
can deliver high flow. It does not mean the battery is full. You must check the battery charge separately. See Section 16.
STEP 4 - DON THE RESPIRATOR WITH FULL FACE MASK
Donning is best done with a partner that can help check adjustments and fit!
If you have long hair, it is recommended you tie the hair back so that it does not interfere with the Full
face harness and the mask seal on the face. To don the Full Face Mask, follow the steps below.
Step 1: Loosen all five straps on
the mask harness to the fullest
extent. Hold the mask in one
hand while you use the other
hand to pull the harness back and
away from the mask.
18