17. Periodic Maintenance & Checking
Recalibration
The respirator can be re-calibrated (see Section 10).
You must recalibrate the internal pressure sensor any time that your CleanSpace Respirator is
exposed to SIGNIFICANT CHANGES in temperature of more than 36°F (20°C). ). It is best
practice to also re-calibrate if the unit has been in storage, particularly if the storage temperature is

18. Approvals

NOTIFIED BODY
Respiratory Protection:
BSI (0086)
Davy Avenue
Knowelhill, Milton Keynes, MK5 8PP
United Kingdom

19. Specifications

CleanSpace Respirators are breath-responsive. Maximum airflow to mask 120 – 220 litre
Air Flow
per minute, depending on filter loading, altitude and battery condition.
Operating:
Operating Temperature Range: -10°C to 45°C. The respirator will stop if temperature of
Temperatures
the unit is greater than 60°C.
Humidity Range
Humidity Range: -10°C Zero to 90%, non-condensing
Operating:
Approximately sea level to 1770m
Altitude Range
Charging:
Charging Temperature Range: 0°C to 35°C
Temperatures
The battery will not accept charge outside of this range.
Storage
Short term storage under 30 days: -10°C to 35°C. (30% - 50% RH)
Conditions
Long term storage (> 30 days): 18°C to 28°C (30% - 50% RH)
Out of direct sunlight, in a clean, dry environment
Shelf Life
Respirator, mask, head harness and neck support: 3 years (under recommended storage
conditions)
HEPA Filter: 3 years
Battery Type
Lithium Ion Polymer battery
Input: 100 – 240VAC, 50 – 60 Hz
Battery Charger
Run Time
Up to 8 hours of operating time.
Operating time is impacted by work rate, mask leak, filter dust loading, altitude and other
factors. Run times may vary widely. IMPORTANT: Short run times (under 4 hours) may be
a sign of poor mask fit – seek advice from a trained safety specialist or CleanSpace
Support www.cleanspacehealth.com
Recharge Time
2 hours (to 95%)
Low Battery
Triggers when the Battery Charge is low and run time is around 5 -15 minutes depending
Alarm
on wok load and filter loading. Audible Battery alarm: 3 beeps, repeated every second,
75dB(A) at ear.
Filter Blocked
Triggers when the filter loading is high and the filter requires replacement.
Alarm
Audible alarm: 2 beeps repeated every second, 75dB(A) at ear. Plus red filter LED
flashes.
Weight
Complete system: 533g (Respirator (CS3001), Filter, Mask, Neck Support, head harness)
Intrinsic Safety
CleanSpace HALO is not an intrinsically safe system
not known.
28