Always correctly use and maintain the internal lithium ion battery packs. Failure to do so may result in fire or explosion or could adversely affect
- Do not charge the internal battery with unapproved chargers or in enclosed cabinets without ventilation or near flammable liquids or gasses, or near
sources of high heat
- Do not immerse the device in water
- Do not use, charge or store the device outside the recommended temperature limits
If the battery alarm sounds (3 beeps, repeated every second), leave the contaminated area immediately and re-charge the battery.
Operating the respirator after the low battery alarm has sounded can cause the flow to fall below the manufacturer's minimum designed flow, which
The Flow Test is not a battery charge test. Three LEDs means that, as of the moment it is tested, the unit can deliver high flow. It does not mean the
IN THE EXTREMELY RARE CIRCUMSTANCE THAT THE BATTERY IS DAMAGED AND ELECTROLYTE COMES IN CONTACT WITH EYES,
WARNING: NEVER DRY THE MASK OR EXHALATION VALVE WITH A CLOTH THAT MAY LEAVE BEHIND LINT.
Lint contamination of the exhalation valve may cause it to leak, resulting in overexposure to contaminants and lead to sickness or death.
WARNING: The Respirator contains a battery, sensitive electronics and a motor. NEVER immerse it in water or use anything wetter than a damp cloth to
WARNING: NEVER CLEAN THE FILTER WITH WATER, COMPRESSED AIR OR A BRUSH. FILTERS ARE DAMAGED BY THE USE OF WATER
Misuse of CleanSpace HALO by cleaning the filter may result in overexposure to contaminants and lead to sickness or death.
You must recalibrate the internal pressure sensor any time that your CleanSpace Respirator is exposed to changes in temperature of more than 20°C

5. Special or Critical User Instructions

This respirator contains a system for synchronising with your breathing and regulating mask pressure. This system
requires re-calibration whenever it experiences a change in temperature of more than 20°C. It is best
practice to also re-calibrate if the unit has been in storage, particularly if the storage temperature is not
known. For the most accurate battery charge indication, the respirator must be plugged into its charger. To be sure
it is fully charged, plug in the charger and wait for all 3 LEDs to light solidly without flashing. If the 3
flashing rapidly, the battery is 95% charged.

6. Limitations of Use

Use this Respirator strictly in accordance with all instructions in these user instructions. Never modify or alter this
product.
Do not remove the respirator until you have left the contaminated area
immediately and sudden onset health reasons to do so (for example you are experiencing dizziness and believe
removing the respirator while you leave the contaminated area may help).
CleanSpace HALO is a tight fitting respirator and National standards recommend the user be clean shaven. If a
satisfactory fit factor cannot be achieved with any of the masks, CleanSpace Respirators must not be used
This Respirator WITH THE HALF MASK DOES NOT PROTECT YOUR EYES. In conditions that may damage
or irritate eyes, use protective eyewear.
Only use your respirator with the parts and accessories listed on the Approval Label.
Do not use the respirator unless it is powered and running normally.
Do not use the respirator while it is being charged via the Battery Charger
Do not use in airborne contaminant concentrations above those specified in your national regulations
Do not use for respiratory protection against unknown atmospheric contaminants or when concentrations of
contaminants are unknown or immediately dangerous to life or health (IDLH).
Do not use in oxygen deficient or oxygen enriched atmospheres. Do not use in flammable or explosive
environments.
Only for use by trained personnel.
respirator performance and result in injury, sickness or death.
may result in overexposure to contaminants and lead to sickness or death.
battery is full. You must check the battery condition separately. See Section 16.
FLUSH WITH WATER IMMEDIATELY AND SEEK URGENT MEDICAL ATTENTION
COMPRESSED AIR OR BY BRUSHING.
clean it.