Check the FILTER carefully. The foam seal must be clean and free from damage of any kind. Examine the
visible internal surfaces for any sign that dust has penetrated the media. If dust is found, do not use the filter.
The body of the filter must be free of cracks or signs of damage. If any sign of impact or scratching is found,
discard the filter. Fit the filter to the Respirator (see Changing the Filter, below)
FILTERS ARE DAMAGED BY THE USE OF COMPRESSED AIR OR BY BRUSHING.
Misuse of CleanSpace Respirator by cleaning the filter may result in overexposure to
Check the MASK to ensure that there are no cracks, tears or dirt; check the mask is not distorted. Check the
mask exhalation valve for damage or dirt build up. If it is dirty, remove the exhalation valve cover. Remove any
dirt, hairs or anything that could affect the seal of the valve against its seat. Check that the valve seat is clean.
Reinstall the valve cover. If the valve is damaged, replace the mask.
Check the HEAD HARNESS is intact and has good elasticity and can be fitted to the attachments on the
mask. It must be adjusted to support the mask to seal comfortably to the face. DO NOT overtighten.
STEP 2 – CALIBRATE
This respirator contains a system for synchronising with your breathing and regulating mask pressure. The system
should be re-calibrated if it experiences a significant change in temperature of more than 20°C. It is best
practice to re-calibrate if the unit has been in storage when the storage temperature is unknown.
Steps to re-calibrate:
1.
Remove the filter from the respirator.
With CleanSpace Respirator in Standby Mode (ie one or more green LEDs lit), press the "Power" and "Test
2.
Flow" buttons on the keypad at the same time.
3.
The red and blue LEDs will both light, and the green battery LEDs will cycle.
Hold the respirator still until the red and blue LEDs turn off (10 – 15 seconds). After this, calibration is complete.
4.
5.
At the end of the calibration process the green LEDs return to indicating battery charge status.
STEP 3 – TEST FLOW RATE
This test checks that the Respirator is able to deliver the Manufacturer's Minimum Design Flow of 120 litres/minute.
1.
Remove the mask from the respirator. Place the respirator flat on a table or other support.
Press and release the button marked "Flow Test"
2.
3.
The respirator will automatically run the Flow Test. Note: During the test, the motor will run fast and air will
blow from the left hand bellows.
4.
After 2 seconds the respirator reports the Test result using the LEDs on the keypad. Use the table below to
interpret the LEDs.
LIGHTS
Meaning
3 LEDs:
PASS (Excellent: flow >180 l/min)**
2 LEDs:
PASS (Good)**
PASS (Acceptable)
1 LED:
ALL LEDs
FAIL (Flow <120 l/min)
FLASH
Do not use the respirator until a new filter has been fitted and / or battery fully charged and the test has
been repeated with a PASS result. Reset the respirator by pressing the Power button. Fully charge the
battery and / or replace the filter. Repeat the flow test. If the respirator fails the test filter is new and battery fully
charged but, contact CleanSpace Technology and do not use until it has been evaluated
** WARNING: The Test Flow Rate is not a battery charge test. Three LEDs means that, as of the moment it is tested, the unit
can deliver high flow. It does not mean the battery is full. You must check the battery charge separately. See Section 16.
STEP 4 - DON THE RESPIRATOR
NOTE – please read STEP 4 all the way through before beginning to don the respirator
Locate your respirator, mask, neck support and harness. Make sure the mask is the size that you used for your
last successful quantitative fit test. This size should be recorded in this manual (see Section 9). ENSURE A
CLEAN FILTER IS FITTED TO THE POWER UNIT.
DO NOT USE COMPRESSED AIR OR A BRUSH TO CLEAN THE FILTER!
contaminants and lead to sickness or death