Changing the basic settings

NOTICE:
If there has been no activity in this menu for approx. 60 seconds, the
menu will automatically close and any changes you have made will not
be applied.
14

Resetting the peak values

The peak values of the saturation level plus temperature can be set back.
In the extended functions menu, please press the key "" until "rSHL" appears.
Press the "mode" key.
Press "" to select "YES" and confirm by pressing the "mode" key. The min/max
temperature values will now be reset.
15

Program enable

15.1 General overview

The instrument has a program enable which must be set to change the settings. The
program enable can be set or cancelled during operation. It provides protection against
unintentional alterations of settings.

15.2 Change program enable

In the extended functions menu, please press the key "" until "PRG" appears.
Press the "mode" key.
You can choose between free programming "FREE" and locked programming "Loc". By
pressing "" or "" you can switch between these options.
Confirm by pressing the "mode" key.
Edition: 2020-04-20
HYDAC ELECTRONIC GMBH
AS 3000 IO-Link
Part No.: 670006 E