11. Specifications
Input signals
Pt1000 (2-/4-wire)
Pt1000 (3-wire)
1) The indication of the measuring error applies to ambient temperatures of 20 ... 40 °C and the measuring time
of 1 second.
2) The indications for Pt100 3-/4-wire apply at a max. lead resistance of 10 Ω.
Inputs for sensor calibration
Input signals
Current signals
Voltage signals
Resistance (2-, 3-,
or 4-wire)
1) The indication of the measuring error applies to ambient temperatures 20 ... 40 °C and the measuring time of
1 second.
WIKA operating instructions digital indicator, model DI35-M
Measuring
span
-200 ... +850 °C
-200 ... +850 °C
Measuring span
0 ... 2 mA
0 ... 5 mA
0 ... 20 mA
4 ... 20 mA
DC -18 ... +18 mV
DC -35 ... +35 mV
DC -75 ... +75 mV
DC -150 ... +150 mV
DC -300 ... +300 mV
DC -500 ... +600 mV
DC -500 ... +1,250 mV
DC -500 ... +2,500 mV
DC -1 ... +5 V
DC -1 ... +10 V
0 Ω ... 100 Ω
0 Ω ... 1 kΩ
0 Ω ... 10 kΩ
Measuring error
in % of the
measuring span
1)
≤ ±0.04 % ±1 digit
≤ ±0.04 % ±1 digit
Measuring error
in % of the
measuring span
1)
≤ ±0.02 % ±1 digit
≤ ±0.02 % ±1 digit
≤ ±0.02 % ±1 digit
≤ ±0.02 % ±1 digit
≤ ±0.06 % ±1 digit
≤ ±0.06 % ±1 digit
≤ ±0.04 % ±1 digit
≤ ±0.03 % ±1 digit
≤ ±0.03 % ±1 digit
≤ ±0.03 % ±1 digit
≤ ±0.03 % ±1 digit
≤ ±0.03 % ±1 digit
≤ ±0.02 % ±1 digit
≤ ±0.01 % ±1 digit
≤ ±0.04 % ±1 digit
≤ ±0.04 % ±1 digit
≤ ±0.04 % ±1 digit
Minimum measuring
time
0.06 s
0.04 s
Minimum
measuring time
0.02 s
0.02 s
0.04 s
EN
53