Herunterladen Inhalt Inhalt Diese Seite drucken

Measuring Time; Display Mode - WIKA DI35-M Betriebsanleitung

Vorschau ausblenden Andere Handbücher für DI35-M:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

7. Description of program numbers

7.2.3 Measuring time

The measuring time determines how often the analogue and switching outputs are
updated. With strongly fluctuating input signals, it is recommended that the measuring
time is adjusted (measuring time > 0.2 ... 1s).
PN
Function
14
Measuring time
The updating of the analogue and switching outputs is made cyclically
with the set measuring time. If the measuring time is set very short, this
can, with an unsteady input signal, lead to jumps with analogue outputs
or to short switching of the switching outputs. The MIN/MAX memory
receives the values on the basis of the set measuring time. Should
the peaks in an unsteady input signal need to be determined, a short
measuring time can be useful.

7.2.4 Display mode

The display mode determines which value is shown on the digital indicator.
PN
Function
15
Display mode
WIKA operating instructions digital indicator, model DI35-M
Parameter
Current, voltage
(PN0 = 1 ... 12 or 33 ... 45)
0.02 ... 10
Resistance (4-/2-wire)
0.04 ... 10
Resistance (3-wire)
0.06 ... 10
Thermocouple
(PN0 = 21 ... 29)
0.04 ... 10
Parameter
1 = instantaneous value
2 = MIN value
3 = MAX value
4 = HOLD value
5 = running average
6 = totaliser value
7 = absolute value
8 = trigger mode
EN
29

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis