Modbus address tables
8.1.35

Controller parameter alarm delay

Modbus PDU address
Hex.
1461
1462
1463
1464
1465
1466
1467
1468
a
Unit: seconds
8.1.36

Controller parameter setpoint values

CAUTION!
Write operations involving many R/W parameters are associated with
saving in an EEPROM or flash memory. These memory modules have only
a limited number of write cycles (approx. 10 000 or 100 000).
Frequent writing of variables can thus result in a memory error in the event
of a power failure.
Fast write cycles should thus be avoided or performed with the aid of the
'external analog inputs' . 'External analog inputs' are not saved in an
EEPROM or flash memory, and are ideal for fast write cycles.
Modbus PDU address
Hex.
1471
1473
1475
1477
1479
147B
147D
147F
a
The unit for the values is based on the configuration of the analog value source selected
for the controller actual value input.
417102397 Rev. 1-10.2019
Data type
Dec.
5217
word
5218
word
5219
word
5220
word
5221
word
5222
word
5223
word
5224
word
Data type
Dec.
5233
float
5235
float
5237
float
5239
float
5241
float
5243
float
5245
float
5247
float
66
Access
Data
r/w
Controller 1
parameter block 1
r/w
Controller 1
parameter block 2
r/w
Controller 2
parameter block 1
r/w
Controller 2
parameter block 2
r/w
Controller 3
parameter block 1
r/w
Controller 3
parameter block 2
r/w
Controller 4
parameter block 1
r/w
Controller 4
parameter block 2
Access
Data
r/w
Setpoint value 1
controller 1
r/w
Setpoint value 1
controller 2
r/w
Setpoint value 1
controller 3
r/w
Setpoint value 1
controller 4
r/w
Setpoint value 2
controller 1
r/w
Setpoint value 2
controller 2
r/w
Setpoint value 2
controller 3
r/w
Setpoint value 2
controller 4
a
a