en

Mounting and installation

3.4

Intended use

The instrument is intended to continuously monitor ambient
air for toxic or combustible gas concentrations. When the gas
concentration exceeds the defined alarm threshold, the built-
in alarm devices (horn and strobes) of the instrument are
triggered. External alarm devices can also be connected and
triggered by the built-in relays. An integration in an alarm
management system is possible using the 4-20mA interface
or relays.
The instrument is not suitable for operation in hazardous
areas with potentially explosive atmospheres.
3.5

Approvals

3.5.1 USA and Canada

Electromagnetic compatibility: Class-B device, residential use
compliant with ICES-3 (B)/NMB-3 (B)
NOTICE
Changes or modifications not expressly approved by the
party responsible for compliance could void the user's
authority to operate the equipment.
This device complies with part 15 of the FCC Rules and
Industry Canada licence-exempt RSS standard(s).
Operation is subject to the following two conditions:
– This device may not cause harmful interference, and
– This device must accept any interference received,
including interference that may cause undesired
operation.
Class B digital device pursuant to part 15 of the FCC
Rules
This equipment has been tested and found to comply with
the limits for a Class B digital device, pursuant to part 15 of
the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no
guarantee that interference will not occur in a particular
installation.
If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
– Reorient or relocate the receiving antenna.
– Increase the separation between the equipment and
receiver.
– Connect the equipment into an outlet on a circuit different
from that to which the receiver is connected.
– Consult the dealer or an experienced radio/TV technician
for help.
10
FCC responsible party:
Draeger Inc.
7256 S. Sam Houston W. Parkway
Suite 100
Houston, Tx 77085 USA
phone: +1 346-802-6111
e-mail: DIHouston.Approvals@draeger.com
4
Mounting and installation
4.1

Opening and closing the instrument

WARNING
Risk of electric shock
Opening the instrument during operation can result in an
electric shock.
► Disconnect power from the instrument and wait at least 1
minute before continuing.
1. Loosen the screws (2) and open the cover.
2. Ensure the following connections are properly made:
a. Wiring screws are tightened to the correct torque (4.4
... 5.3 Lbs. In. / 0.5 ... 0.6 Nm).
b. All cable connectors are secured with screws.
c. The sensor connector is plugged (PointGard 2x00 CAT
only).
d. Cable cores of cables that lead into the housing are
secured with a cable ties.
3. Close cover and tighten screws (2) with correct torque (22
Lbs. In./ 2.5 Nm).
4.2

Preparing the power supply cable

Power supply cables have to be ordered separately and
connected to the instrument on-site. For recommended cable
properties and part numbers see 16.2 Power supply. Every
cable used has to comply with the properties and has to be
prepared following these instructions.
1. Strip the power supply cable.
2. Adjust the length of the wires (AC version only). The PE
wire must be longer than the other wires.
L/N: 50 mm +10 mm / 1.9 in + 0.4 in
PE: 65 mm + 10 mm / 2.5 in + 0.4 in
3. Connect ferrules. For recommended ferrule properties,
see 16.2 Power supply.