ENDOGATOR
Hybrid Irrigation Tubing
Intended Use:
The ENDOGATOR™ Hybrid Irrigation Tubing (tubing and accessories to accommodate various GI endoscopes and irrigation
pumps) is intended to provide irrigation via sterile water and to supply air (via an air pump) or CO 2 (via a CO 2 supply) along
with sterile water during GI endoscopic procedures when used in conjunction with an irrigation pump or cautery unit.
Warnings and Precautions:
ALWAYS prime the air/water channel prior to inserting the GI endoscope into the patient.
ALWAYS prime the auxiliary water channel prior to inserting the GI endoscope into patient.
NEVER attach or reattach the ENDOGATOR Hybrid Irrigation Tubing to a GI endoscope that is/was inside a patient
and has not been reprocessed.
ALWAYS close the pinch clip on the air/water tube before disconnecting the air/water tubing from the GI endoscope.
NEVER turn off the air processor/CO 2 supply while the ENDOGATOR Hybrid Irrigation Tubing is attached to a
GI endoscope.
DO NOT REMOVE THE BACKFLOW VALVE. The backflow valve is permanently attached to the irrigation tubing.
The backflow valve attached to the ENDOGATOR Hybrid Irrigation Tubing must be in place. If the backflow valve is
missing, then discard the entire ENDOGATOR Hybrid Irrigation Tubing, accessories and water bottle.
DO NOT RESTERILIZE OR REPROCESS.
If any of the above prevention guidelines were not followed, or there is any reason to believe that the ENDOGATOR
Hybrid Irrigation Tubing system was contaminated, immediately discard the entire tubing, accessories and water
bottle. Replace with a new tubing and unopened sterile water bottle.
ALWAYS inspect the GI endoscope's o-rings and gaskets (air/water button, GI endoscope connection, etc.) for damage.
Damaged components can result in less than optimal performance of the ENDOGATOR Hybrid Irrigation Tubing.
Sterility is not guaranteed if package has been opened or damaged. Do not use if packaging or product is damaged.
ENDOGATOR Hybrid Irrigation Tubing is for 24 hour use (discard daily).
Contact customer service for assistance.
This product contains DEHP (Bis(2-ethylexyl)phthalate) and may be harmful to sensitive populations (children and
pregnant/nursing mothers).
Due to the short duration of use and the slow migration rate of DEHP from leachable materials into aqueous fluid,
the residual chemical safety risk is deemed negligible. The risk and benefits of using this device shall be carefully
evaluated by a clinician on a case-by-case basis.
Federal (USA) law restricts the sale of this device by or on the order of a physician.
PRECAUTION: If sterile water bottle needs replacing during a procedure, refrain from depressing the air/water valve.
There is a potential for backflow to occur if the air/water valve is depressed while no sterile water bottle is connected
to the tubing. If this occurs, replace the tubing immediately.
STERILIZED USING
ETHYLENE OXIDE
ENDOGATOR™ Hybrid Irrigation Tubing for OLYMPUS
Series GI Endoscopes with ENDO STRATUS™ Pump, OLYMPUS OFP
or BOSTON SCIENTIFIC
ENDOGATOR Hybrid Irrigation Tubing for OLYMPUS 140/160/180
Series GI Endoscopes with OLYMPUS AFU-100, OLYMPUS OFP,
24 HOUR USE
(DISCARD DAILY)
EN
#100609 (includes "S" series)
140/160/180
®
®
ENDOSTAT™ II Pump Irrigation Units
#100610 (includes "S" series)
ERBE
EIP2 or EGP-100 Irrigation Units
®
CONTAINS DEHP
(Bis (2-ethylhexyl) phthalate)

Werbung

Inhaltsverzeichnis
loading

Inhaltszusammenfassung für Cantel Medical Medivators ENDOGATOR S Serie

  • Seite 1 ENDOGATOR ™ Hybrid Irrigation Tubing #100609 (includes “S” series) ENDOGATOR™ Hybrid Irrigation Tubing for OLYMPUS 140/160/180 ® Series GI Endoscopes with ENDO STRATUS™ Pump, OLYMPUS OFP or BOSTON SCIENTIFIC ® ENDOSTAT™ II Pump Irrigation Units #100610 (includes “S” series) ENDOGATOR Hybrid Irrigation Tubing for OLYMPUS 140/160/180 Series GI Endoscopes with OLYMPUS AFU-100, OLYMPUS OFP, ERBE EIP2 or EGP-100 Irrigation Units...
  • Seite 2 Instructions for Use (by product number): 1. Open the sterile peel pack and remove the ENDOGATOR™ Hybrid Irrigation Tubing. Open a bottle of sterile water (1000cc or 1500cc). Drop the angle cut tubing bundle into the sterile water bottle and firmly tighten the bottle cap to ensure a secure seal.
  • Seite 3: Warnhinweise Und Vorsichtsmaßnahmen

    ENDOGATOR ™ Hybrid Irrigation Tubing Nr. 100609 (inkl. „S“-Serie) ENDOGATOR™ Hybrid-Spülschlauch für OLYMPUS GI-Endoskope ® der Serien 140/160/180 mit ENDO STRATUS™-Pumpe, OLYMPUS OFP- oder BOSTON SCIENTIFIC ® ENDOSTAT™ II Pumpenspüleinheiten Nr. 100610 (inkl. „S“-Serie) ENDOGATOR Hybrid-Spülschlauch für OLYMPUS GI-Endoskope der Serie 140/160/180 mit OLYMPUS AFU-100, OLYMPUS OFP, ERBE EIP2 oder EGP-100 Spüleinheiten ®...
  • Seite 4 Gebrauchsanweisung: 1. Öffnen Sie die sterile Peel-Verpackung und entnehmen Sie den ENDOGATOR™ Hybrid-Spülschlauch. Öffnen Sie eine Flasche mit sterilem Wasser (1000 ml oder 1500 ml). Lassen Sie das schräg angeschnittene Schlauchbündel in die Sterilwasserflasche sinken und schrauben Sie den Deckel dicht zu. Lösen Sie den 24-Stunden-Aufkleber von der Verpackung und bringen Sie ihn an der Sterilwasserflasche an.
  • Seite 5: Avertissements Et Précautions

    ENDOGATOR ™ Hybrid Irrigation Tubing n° 100609 (incluant la série « S ») Tubulure d'irrigation hybride ENDOGATOR™ pour endoscopes GI OLYMPUS série 140/160/180 avec pompe ENDO STRATUS™, ® unités d'irrigation à pompe OLYMPUS OFP ou BOSTON SCIENTIFIC ENDOSTAT™ II ® n° 100610 (incluant la série « S ») Tubulure d'irrigation hybride ENDOGATOR pour endoscopes GI OLYMPUS série 140/160/180 avec les pompes d'irrigation OLYMPUS AFU-100, OLYMPUS OFP, ERBE...
  • Seite 6 Mode d'emploi 1. Ouvrir l'emballage stérile à pellicule pelable et sortir la tubulure d'irrigation hybride ENDOGATOR™. Ouvrir un flacon d'eau stérile (1 000 ou 1 500 ml). Introduire l'ensemble des tubulures coupées en biseau dans le flacon d'eau stérile et resserrer fermement le bouchon du flacon afin d'en garantir l'étanchéité. Sortir l'étiquette « 24 h » de la poche puis l'apposer sur le flacon d'eau stérile.
  • Seite 7: Advertencias Y Precauciones

    ENDOGATOR ™ Hybrid Irrigation Tubing N.º 100609 (incluye serie “S”) Tubo de irrigación híbrido ENDOGATOR™ para gastroscopios OLYMPUS de las series 140/160/180 con bomba ® ENDO STRATUS™ y unidades de irrigación de bomba OLYMPUS OFP o BOSTON SCIENTIFIC ENDOSTAT™ II ®...
  • Seite 8: Instrucciones De Uso

    Instrucciones de uso: 1. Abra el envase desprendible estéril y extraiga el tubo de irrigación híbrido ENDOGATOR™. Abra una botella de agua estéril (de 1.000 o 1.500 ml). Deje caer el haz de tubo cortado en ángulo en el interior de la botella de agua estéril y apriete firmemente la tapa de la botella para asegurar un cierre hermético.
  • Seite 9: Uso Previsto

    ENDOGATOR ™ Hybrid Irrigation Tubing #100609 (inclusa serie “S”) Sistema ibrido di irrigazione ENDOGATOR™ per endoscopi GI OLYMPUS serie 140/160/180 con pompa ENDO STRATUS™, ® OLYMPUS OFP o unità di irrigazione con pompa BOSTON SCIENTIFIC ENDOSTAT™ II ® #100610 (include la serie “S”) Sistema ibrido di irrigazione ENDOGATOR per endoscopi GI OLYMPUS serie 140/160/180 con pompe di irrigazione OLYMPUS AFU-100, OLYMPUS OFP, ERBE...
  • Seite 10: Istruzioni Per L'uso

    Istruzioni per l’uso 1. Aprire la confezione sterile peel-pack ed estrarre il sistema ibrido di irrigazione ENDOGATOR™. Aprire una bottiglia di acqua sterile (da 1000 o 1500 cc). Immergere l’estremità a taglio obliquo del fascio di tubi nella bottiglia di acqua sterile e stringere bene il tappo della bottiglia per assicurare la tenuta.
  • Seite 11: Beoogd Gebruik

    ENDOGATOR ™ Hybrid Irrigation Tubing #100609 (inclusief S-serie) ENDOGATOR™ hybride irrigatieslang voor OLYMPUS 140/160/180-serie ® GI-endoscopen met ENDO STRATUS™-pomp, OLYMPUS OFP of BOSTON SCIENTIFIC ® ENDOSTAT™ II pompirrigatie-units #100610 (inclusief S-serie) ENDOGATOR hybride irrigatieslang voor OLYMPUS 140/160/180-serie GI-endoscopen met OLYMPUS AFU-100, OLYMPUS OFP, ERBE EIP2 of EGP-100-irrigatie-units ®...
  • Seite 12 Gebruiksinstructies 1. Open de steriele peelverpakking en haal de ENDOGATOR™ hybride irrigatieslang eruit. Open een fles steriel water (1000 cc of 1500 cc). Laat de gehoekt afgesneden slangbundel in de fles steriel water zakken en draai de dop stevig op de fles om voor goede afdichting te zorgen. Haal het 24 uur-etiket van de zak en plak het op de fles steriel water. Schrijf de datum en de tijd op het etiket.
  • Seite 13: Beregnet Brug

    ENDOGATOR ™ Hybrid Irrigation Tubing nr. 100609 (inklusive “S”-serien) ENDOGATOR™-hybridskylleslange til OLYMPUS serie 140/160/180 ® GI-endoskoper med ENDO STRATUS™-pumpe, OLYMPUS OFP- eller BOSTON SCIENTIFIC ® ENDOSTAT™ II-skyllepumpeenheder nr. 100610 (inklusive “S”-serien) ENDOGATOR-hybridskylleslange til OLYMPUS serie 140/160/180 GI-endoskoper med OLYMPUS AFU-100-, OLYMPUS OFP-, ERBE EIP2- eller EGP-100-skylleenheder ®...
  • Seite 14 Brugsanvisning: Åbn den sterile peel-pakke, og tag ENDOGATOR™-hybridskylleslangen ud. Åbn en flaske med sterilt vand (1000 ml eller 1500 ml). Sæt det vinkelskårne slangebundt ind i flasken med sterilt vand, og skru låget fast, så flasken er sikkert forseglet. Tag 24-timersetiketten ud af posen, og sæt den på flasken med sterilt vand. Skriv dato og klokkeslæt på etiketten. BEMÆRK: Hvis CO 2 -kilden anvendes, forbindes CO 2 -indgangen med tilbageløbsventilen på...
  • Seite 15: Advarsler Og Forsiktighetsregler

    ENDOGATOR ™ Hybrid Irrigation Tubing nr. 100609 (inkludert “S”-serien) ENDOGATOR™ hybridskylleslange for OLYMPUS 140/160/180-serien ® av GI-endoskoper med ENDO STRATUS™-pumpe, OLYMPUS OFP- eller BOSTON SCIENTIFIC ® ENDOSTAT™ II pumpeskylleenheter nr. 100610 (inkludert “S”-serien) ENDOGATOR hybridskylleslange for OLYMPUS 140/160/180-serien av GI-endoskoper med OLYMPUS AFU-100, OLYMPUS OFP, ERBE EIP2 eller EGP-100 skylleenheter ®...
  • Seite 16 Bruksanvisning: 1. Riv opp den sterile pakningen, og fjern ENDOGATOR™-hybridskylleslangen. Åpne en flaske med sterilt vann (1000 ml eller 1500 ml). Slipp den vinkelkuttede slangebunten ned i flasken med sterilt vann, og skru flaskekorken godt igjen slik at den er helt tett. Ta 24-timersetiketten ut av posen, og fest den på flasken med sterilt vann. Skriv datoen og klokkeslettet på...
  • Seite 17: Utilização Prevista

    ENDOGATOR ™ Hybrid Irrigation Tubing #100609 (inclui a série “S”) Tubo de irrigação híbrido ENDOGATOR™ para endoscópios gastrintestinais OLYMPUS da série 140/160/180 com bomba ® ENDO STRATUS™, bombas de irrigação OLYMPUS OFP ou BOSTON SCIENTIFIC ENDOSTAT™ II ® #100610 (inclui a série “S”) Tubo de irrigação híbrido ENDOGATOR para endoscópios gastrintestinais OLYMPUS da série 140/160/180 com unidades de irrigação OLYMPUS AFU-100,...
  • Seite 18: Instruções De Utilização

    Instruções de utilização: 1. Abra a embalagem de abertura fácil e retire o tubo de irrigação híbrido ENDOGATOR™. Abra uma garrafa de água esterilizada (1000cc ou 1500cc). Coloque o agrupamento de tubos de ângulo cortado dentro da garrafa de água esterilizada e aperte firmemente a tampa da garrafa para garantir que está...
  • Seite 19: Avsedd Användning

    ENDOGATOR ™ Hybrid Irrigation Tubing nr. 100609 (inkluderar “S”-serien) ENDOGATOR™ hybridspolslang för magtarmendoskop i OLYMPUS ® serierna 140/160/180 med ENDO STRATUS™ pump, OLYMPUS OFP eller BOSTON SCIENTIFIC ® ENDOSTAT™ II pumpspolenheter nr. 100610 (inkluderar “S”-serien) ENDOGATOR hybridspolslang för magtarmendoskop i OLYMPUS-serierna 140/160/180 med OLYMPUS AFU-100, OLYMPUS OFP, ERBE EIP2 eller EGP-100 spolenheter...
  • Seite 20 Bruksanvisning: 1. Öppna den sterila rivförpackningen och ta ut ENDOGATOR™ hybridspolslang. Öppna en steril vattenflaska (1000 ml eller 1500 ml). Släpp ner den vinkelskurna slangbunten i den sterila vattenflaskan och dra fast flaskans lock stadigt tills det är tättslutande. Ta bort 24-timmarsetiketten från påsen och sätt på etiketten på den sterila vattenflaskan. Skriv datum och tid på...
  • Seite 21: Предупреждения И Меры Предосторожности

    ENDOGATOR ™ Hybrid Irrigation Tubing № 100609 (в том числе серия «S»)  Гибридная ирригационная трубка ENDOGATOR™ для ЖКТ эндоскопов OLYMPUS серии 140/160/180 с насосом ENDO STRATUS™, ® а также с ирригационными насосными установками OLYMPUS OFP или BOSTON SCIENTIFIC ENDOSTAT™ II ® № 100610 (в том числе серия «S»)  Гибридная ирригационная трубка ENDOGATOR для ЖКТ эндоскопов...
  • Seite 22: Инструкция По Применению

    Инструкция по применению: 1. Вскройте стерильную упаковку и извлеките гибридную ирригационную трубку ENDOGATOR™. Откройте флакон со стерильной водой (1000 куб.см или 1500 куб.см). Опустите пучок срезанных под углом трубок во флакон со стерильной водой и плотно закрутите крышку флакона для обеспечения герметичности. Отклейте 24-часовую этикетку...
  • Seite 23 ENDOGATOR ™ Hybrid Irrigation Tubing Nº 100609 (incluye la Serie “S”) Tubo de irrigación híbrido ENDOGATOR™ para endoscopios OLYMPUS Serie GI 140/160/180 bomba ENDO STRATUS™ y bombas ® de irrigación OLYMPUS OFP o BOSTON SCIENTIFIC ® ENDOSTAT™ II Nº 100610 (incluye la Serie “S”) Tubo de irrigación híbrido ENDOGATOR para endoscopios OLYMPUS Serie GI 140/160/180 con AFU-100 OLYMPUS, OFP OLYMPUS, ERBE...
  • Seite 24 Instrucciones de uso: 1. Abra el paquete removible estéril y extraiga el tubo de irrigación híbrido ENDOGATOR™. Abra la botella de agua estéril (1000 cc o 1500 cc). Coloque el sistema del tubo de corte angular en la botella de agua estéril y ajuste con firmeza la tapa de la botella para que quede cerrada en forma hermética.
  • Seite 25: Uso Pretendido

    ENDOGATOR ™ Hybrid Irrigation Tubing #100609 (inclui a série “S”) Tubos de irrigação híbridos ENDOGATOR™ para os endoscópios gastrointestinais OLYMPUS série 140/160/180 ® com bomba ENDO STRATUS™, unidades de irrigação por bomba OLYMPUS OFP ou BOSTON SCIENTIFIC ENDOSTAT™ II ® #100610 (inclui a série “S”) Tubos de irrigação híbridos ENDOGATOR para os endoscópios gastrointestinais OLYMPUS série 140/160/180...
  • Seite 26: Instruções De Uso

    Instruções de uso: 1. Abra a embalagem destacável estéril e remova os tubos de irrigação híbridos ENDOGATOR™. Abra um frasco de água estéril (1000 cc ou 1500 cc). Coloque o conjunto de tubos dobrado no frasco de água estéril e feche firmemente a tampa do frasco para garantir uma vedação segura.
  • Seite 28 ENDO STRATUS™ and ENDOGATOR™ are trademarks of Medivators Inc. All other company or product names referenced in this document are trademarks of their respective owner. Manufactured in the USA by: Medivators Medivators Inc. Medivators BV Cantel (Australia) Pty Ltd 14605 28th Avenue North Sourethweg 11 300 Lorimer Street, Port Melbourne, Victoria 3207, Australia 3150 Pollok Drive...

Inhaltsverzeichnis