2.6.2

Installation site

WARNING!
Overheating from inadequate cooling!
The device might be damaged!
Maintain the minimum distances from the walls! Dimen-
sions are given on page 33.
Do not change, obstruct or block the air inflow and
outflow to the device or make any unauthorized
changes to the device.
Environmental conditions
Optimum ambient temperature: Between 20 °C and 35 °C
Storage temperature: Between -15 °C and +60 °C (the well, pump and
the hoses must not contain any water)
Relative humidity: Between 30 % and 75 %
Air pressure: Between 700 hPa and 1060 hPa
Water and cooling conditions
Permissible water pressure (water inlet): 8 bar
Maximum coolant temperature: 16 °C
The device is not to be used outdoors.
Suitable connectors for the power and the water supplies must be available,
see Technical data on page 32.
A conversion is required in order to operate the device without problems at
altitudes exceeding 2000 meters above sea level. Please contact the cus-
tomer service department (see page 2) without fail before putting the device
into operation.
2.7

Commissioning

Initial commissioning will be undertaken by the manufacturer's own service
personnel or by an authorized service company. The device will be handed
over ready for use.
If the device has not been used for a longer period of time then it must first be
inspected by our customer service staff or by another authorized company
before being put back into operation again.
The device is only to be used if it is working correctly!
None of the safety devices (e.g. switch) or safety warnings are to be removed
or deactivated as this would impair the safe operation of the device!
The device must never be used if it is empty.
A water cooling system must be connected up to the device if it will be used
for more than 3 hours per day. The coolant hose must never be removed
once it has been connected up.
We recommend that a water shut-off valve is used (water flow controller).
High frequency mobile communication systems can interfere with the way the
device functions.
Information for the operator
Operating instructions – 9