Sterilization
Note: extra equipment is needed for this function.
Select Menu 11, and push
Push
to start the process, which lasts 20 minutes. If the door is opened, the process will be
START
STOP
interrupted. Push
Fan speed
The evaporator fans are normally set at max. speed (100%). However it is possible to reduce the
speed to 50%. By pushing
be changed by pushing
Speed
100%
Push
to exit the programme.
Cleaning
Always disconnect the cabinet before cleaning.
Do not flush compressor compartment and evaporator with water as this may cause short-circuits in
the electrical system.
The cabinet should be cleaned internally with a mild soap solution at suitable intervals and checked
thoroughly before it is put into operation again.
For the external maintenance – use stainless steel polish.
Cleansing agents containing chlorine or compounds of chlorine as well as other corrosive
means, may not be used, as they might cause corrosion to the stainless panels of the cabinet and the
evaporator system.
The compressor compartment and in particular the condenser must be kept free from dust and dirt.
This is best done with a vacuum cleaner and a brush.
765041088
to enter the menu:
to stop sterilization. Push
during the process, the current fan speed is displayed. The speed can
.
Start Steriliz
No
MENU
to exit the programme.
Speed
50%
Ok
32