7.2
Warning messages during operation
Warning code and
Prior-
ID letter
ity
Warning: 1
Q Simulation
Warning: 2
Totalizer reset
Warning: 3
Test Mode
Warning: 4
Function test
1)
Warning: 7
Ext.Dat.loaded
1)
Warning: 8a
Update int.Dat
1)
Warning: 8b
Update ext.Dat.
Warning: 9a
Overflow > F
Warning: 9b
Overflow < R
Warning: 9c
Overflow Diff.
D184B141U03
Description
2
Before operation with flow, a
flowrate can be simulated. The
output values correspond to the
simulated flowrate entered.
1
All totalizers (forward/reverse,
diff. and overflow counters are
reset.)
3
Only for checking the transmitter
using Simulator 55XC4000.
4
If during HART communication,
the functional test for the contact
output or contact input was
started, the Warning 4 is
displayed.
9
Transmitter has recognized
different flowmeter sensor data
and loaded it in the internal
FRAM. Loaded are the system
and flowmeter sensor data.
10
Transmitter has detected an error
in the internal FRAM and has
repaired the data with content of
the external FRAM.
Transmitter has detected an error
in the external FRAM and has
repaired the data with content of
the internal FRAM.
5
Forward totalizer has reached
the maximum count and has
restarted at zero.
6
Reverse totalizer has reached
the maximum count and has
restarted at zero.
7
Diff. totalizer has reached the
maximum count and has
restarted at zero.
Possible cause
Simulation mode on.
External totalizer
reset was performed.
Test Mode on.
Function test on.
Defective FRAM,
repair unit, replace
instrument.
Defective FRAM,
repair unit, replace
instrument.
Defective FRAM or
data was modified.
Max. totalizer
overshot, 1 overflow
as value was
totalized.
Max. totalizer
overshot, 1 overflow
as value was
totalized.
Max. totalizer
overshot, 1 overflow
as value was
totalized.
FSM4000
Messages and tests
Corrective action
After completing the
simulation program,
switch off the
parameter "Simulation
Mode".
Open the switch at the
contact input (terminal
81, 82).
After running the
simulation on the
simulator, switch off the
parameter "Test Mode".
After completing the
self check test, exit the
routine.
Write down all totalizer
values, totalizers
should be reset.
Check the setup data
and correct, if
necessary.
Check the setup data
and correct, if
necessary.
Reset totalizer, if
necessary.
Reset totalizer, if
necessary.
Reset totalizer, if
necessary.
EN - 41