Messages and tests
7.1
Error messages during operation and with data entry
The error messages listed below include explanations of the error codes shown in the display. When entering
information, the error codes 0 ... 6, A, B, C, D, E, G do not appear.
Error code and
Priori
clear text message
ty
Error: 0
Empty pipe
Error: 1
AD Converter/DSP
Error: 2
Driver
Error: 3
Flow > 103%
Error: 4
Zero Return
Error: 5
Database
Error: 6
Totalizer
D184B141U03
Description
5
The meter tube is not
full.
4
The AD converter is
saturated and is not
responding.
7
Positive or negative
reference too small.
6
The max. measuring
range was exceeded
by more than 3%.
8
The flowrate is set to
zero; the totalizer is
stopped.
2
Loss of the internal
database.
9
Error in totalizer > F.
Error totalizer < R
Error totalizer
Possible cause
The pipeline is empty
and the electrodes are
not coming into contact
with the fluid.
Input metering signal is
too large.
The AD converter/DSP is
defective.
Check the cables, no
reference voltage is
present. The current
limiter in the driver has
responded because the
driver current is
insufficient. Defective
driver fuse.
The flowrate is set too
high or the measuring
range is set too small.
The external contact is
closed.
Data memory module is
defective.
The forward totalizer is
damaged.
The reverse totalizer is
damaged.
The totalizer forward,
reverse or diff. totalizer is
damaged.
FSM4000
Messages and tests

Corrective action

Refill the meter tube. Detector
empty pipe is on, but the
adjustment has not been run.
Adjust DEP.
Check ground (flowmeter
sensor). Check the signal cable
and the measuring range
setting; the measuring range
selected may be too low.
Replace the DSP board.
Check connection board and
transmitter.
Increase the measuring range,
reduce the flowrate.
The external contact is open
again.
Turn unit off and on again, call
up and run functional test for
transmitter.
Reset forward/reverse totalizer
or preset new values in
totalizer.
Totalizer forward, reverse
defective.
Check transmitter and wiring.
EN - 39