Messages and tests
Error code and
Prior-
clear text message
ity
Error: A
10, 11 Max alarm limit value. The configured Max alarm
Max. Alarm
Error: B
Min. Alarm
Error: C
Ext. data base
Error: D
Old Primary
Error: E
DC to high
Error: F
FRAM in Primary
Error: G
Error Diagnosis
1)
Replacing a defective FRAM.
If the FRAM is defective and startup cannot be performed, a new FRAM can be requested from the Göttingen
plant. To properly process requests, you must include the ABB order number and unit number of the flowmeter
sensor. After installing the FRAM and switching on the power, you can start up the system. Review all
flowmeter sensor data and setup data for the system, and enter this information again, if necessary.
40 - EN
Description
Min alarm limit value.
3
External database
FRAM is defective or
not available.
12
Flowmeter sensor type
from the older model
flowmeter series was
selected.
13
Increased analog
reset, measurement
signal with large DC
1
Missing data from
external FRAM.
14
Diagnostic or wiring
error
Possible cause
for the flowrate was
exceeded.
The flowrate is below the
configured Min alarm
Missing FRAM or
1)
defective FRAM
.
Configuration under the
parameter "Primary
Setup" is incomplete.
Air bubbles, deposits on
electrodes, too high
interference signal.
Empty pipe.
FRAM for the flowmeter
sensor is still installed in
the terminal box.
Min-Max over/undershoot
for diagnostic values.
Wiring error for electrode
circuit. Coil short-circuit
to ground
FSM4000
Corrective action
Reduce flowrate.
Increase flowrate.
Install and screw in place the
FRAM for the associated
flowmeter sensor to the
connection board in the field
housing unit. See chapter
the FRAM cannot be read, it
must be replaced.
Complete configuration under
"Primary Setup". See also
chapter 11 at the operating
instruction.
Activate "Noise reduction" in
the operating mode submenu.
Use a air separator, clean
electrodes, activate empty pipe
detector.
Contact ABB Service.
Install and screw in place the
FRAM for the associated
flowmeter sensor to the
connection board in the field
housing unit. See chapter
Retrieve details of error
message under the Status
submenu in the Diagnostic error
register. Adjust thresholds, if
necessary.
Retrieve details of error
message under the Status
submenu in the Wiring error
register.
D184B141U03
5 . If
H
5 .
H